Differences
This shows you the differences between two versions of the page.
Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
library:mri_standard_parameters [2025/05/06 15:21] – scott | library:mri_standard_parameters [2025/06/25 14:20] (current) – scott | ||
---|---|---|---|
Line 1: | Line 1: | ||
=====MRI Standard Parameters===== | =====MRI Standard Parameters===== | ||
^Parameter^T1 SE/FSE^T2 FSE^STIR IR-FSE^T2 FLAIR IR-FSE^GRE^DWI^ | ^Parameter^T1 SE/FSE^T2 FSE^STIR IR-FSE^T2 FLAIR IR-FSE^GRE^DWI^ | ||
- | |TR|400-700|3000+ (2000+ if Restore Pulse)|3000+|9000+|VAR|8000+| | + | |TR|400-700|3000+ (1500+ if Restore Pulse)|3000+|9000+|VAR|8000+| |
|TE|MIN; < | |TE|MIN; < | ||
|ETL|3-5 If FSE|16-24|12-16|16-28|N/ | |ETL|3-5 If FSE|16-24|12-16|16-28|N/ | ||
Line 10: | Line 10: | ||
|SLICE THICKNESS|2 - 3mm|2 - 3mm|2 - 3mm|2 - 3mm|2 - 3mm|2 - 5mm| | |SLICE THICKNESS|2 - 3mm|2 - 3mm|2 - 3mm|2 - 3mm|2 - 3mm|2 - 5mm| | ||
|SLICE SPACING|10%|10%|10%|10%|10%|10%| | |SLICE SPACING|10%|10%|10%|10%|10%|10%| | ||
- | |NOTES| -Pre and post contrast \\ -Choose FSE for metal artifacts| -High ETL may cause blurring \\ -If using FatSat, decrease TE to 70-80 -TE 500+ for myelo| \\ -Superior FatSat over large FOV \\ -Do not scan post contrast -Low inherent SNR| \\ -Sensitive to subtle contrast | + | |NOTES| -Pre and post contrast \\ -Choose FSE for metal artifacts| -High ETL may cause blurring \\ -If using FatSat, decrease TE to 70-80 -TE 500+ for myelo| |
Provided by Scott Johnson, MS RT (R)(MR) - edited by meuon, a weird spinning particle in the machine. | Provided by Scott Johnson, MS RT (R)(MR) - edited by meuon, a weird spinning particle in the machine. | ||