- Increase in-plane resolution a bit | - Increase in-plane resolution a bit | ||
- The artifact will smear out in the frequency encoding direction, consider swapping | - The artifact will smear out in the frequency encoding direction, consider swapping | ||
- | - Increase | + | - Increase |
+ | |||
+ | The susceptibility artifact evolves over time throughout the frequency encoding period. Increasing the bandwidth helps by reducing the echo spacing so that the time between data points is shorter. This also reduces the time between strong refocusing pulses in the FSE sequence, further suppressing the artifact. In addition, the amplitude of the frequency encoding gradient is increased, which also helps to reduce the severity of the artifact. Similar ends are achieved by increasing the imaging matrix and reducing the slice thickness; it will impose stronger gradients onto the microchip. Swapping the frequency encoding direction won't suppress the artifact, but it will redirect it or change the shape of it, depending on the orientation of the microchip. All of these parameter changes are likely to increase scan time, but it is best to avoid using parallel imaging if the artifact is significant, | ||
+ | |||
+ | ===I need thinner slices! What can I do?=== | ||
+ | |||
+ | When imaging very small anatomy with 2D FSE sequences, it can be difficult to reduce slice thickness to much less than 2mm without sacrificing too much scan time or SNR. If less than 2mm imaging is required, it is necessary to move to a 3D sequence. Any time this is done, it is a good idea to discuss with the clinician the caveats for whichever sequence is being substituted. Before substituting, | ||
+ | |||
+ | * What is the goal of this sequence? Am I trying to demonstrate contrast enhancement, | ||
+ | * Is true T2 weighting required or is bright fluid contrast good enough? | ||
+ | * What are the drawbacks to this sequence? Are there any particular artifacts, contrast behaviors, or scan time limitations? | ||
+ | * What is the inherent SNR of the sequence? | ||
+ | * What are the specific advantages of this sequence? | ||
+ | |||
+ | Below are some commonly available sequences, with their vendor specific names, where to find them, caveats, and advantages: | ||
+ | |||
+ | ==Balanced Steady State Free Precession with Phase Cycling (FIESTA-C, CISS)== | ||
+ | |||
+ | Where to find: Most vendors have this saved in an IAC protocol, occasionally spine. | ||
+ | |||
+ | Caveats: T2/T1 weighting leads to unusual contrast behaviors; fluid will be very bright, edema won't be well demonstrated, | ||
+ | |||
+ | Advantages: Very high inherent SNR; it's possible to get really high resolution or very thin slices. It can be run isotropically for reformats. Bright fluid contrast can help with anatomic information | ||
+ | |||
+ | ==RF Spoiled Fast Gradient Echo (FLASH, FSPGR)== | ||
+ | |||
+ | Where to find: Most vendors have this saved in a brain protocol | ||
+ | |||
+ | Caveats: Vessel signal will be bright, especially after contrast. These sequences are less sensitive to gadolinium enhancement than FSE. Be aware that susceptibility/ | ||
+ | |||
+ | Advantages: Pretty robust sequences with relatively high inherent SNR. There are usually different fat saturation options available such as SPECIAL/ | ||
+ | |||
+ | ==RF Spoiled Fast Gradient Echo with k space modifications (LAVA, VIBE)== | ||
+ | |||
+ | Where to find: Most vendors have this saved in a routine liver protocol | ||
+ | |||
+ | Caveats: Vessel signal will be bright, especially after contrast. These sequences are less sensitive to gadolinium enhancement than even the FSPGR/ | ||
+ | |||
+ | Advantages: These are really designed to run for speed more than anything else, so they' | ||
+ | |||
+ | ==3D Fast Spin Echo (CUBE, SPACE)== | ||
+ | |||
+ | Where to find: Most vendors have this saved in a spine or brain protocol | ||
+ | |||
+ | Caveats: These sequences have been around for awhile, but only effectively executed more recently on modern scanners with faster gradients, so keep in mind how old the scanner is; often a 3D GRE will be less blurry on older scanners. There are going to be multiple contrasts to choose from, it is best to select one out of the box that is closest to the same body part and contrast that is needed, as modifying parameters to change contrasts may be a lot of work and result in errors. There may be coil combination, | ||
+ | |||
+ | Advantages: Very high SNR, these can be run isotropically or with very thin slices. Due to the long ETL and varying flip angle intravascular signal will be dark, which is excellent for vessel wall imaging | ||
+ | |||
+ | ==Multi Echo GRE (MERGE, MEDIC)== | ||
+ | |||
+ | Where to find: Most vendors have this saved in a cervical spine protocol | ||
+ | |||
+ | Caveats: Relatively lower SNR for 3D sequences, may have a bit of a longer run time. Sensitive to motion artifact, so carefully consider the phase encoding direction when scanning; axial in the thoracic spine is often low image quality. Suggested planes: Cervical Spine: Axial or sagittal Thoracic Spine: Sagittal or Dorsal, Lumbar Spine: Any plane should look ok. isotropic imaging will run very long, so it's not usually worth it since parallel imaging will decrease the already lower SNR. Best to run with slice thickness 1.5mm or greater. May have an FOV limitation depending on software level. | ||
+ | |||
+ | Advantages: Much more robust T2* contrast due to multiple combined echoes at both short and long TE's. Great anatomic contrast for the spine as bone will be black, discs and joints will be very bright, CSF and Fat are different shades of grey, and grey/white matter contrast is excellent. Sensitive to hemorrhage around the spinal cord. Hydrated disc extrusions can be more apparent than on T2. Can have multiple fat saturation options such as SPECIAL, SPAIR, Water Excitation, as well as regular Fatsat. Can be great for a wobbler' | ||
+ | |||
- | There are a few reasons for why all these parameter changes will reduce the susceptibility artifact. |