=====MRI Standard Parameters===== ^Parameter^T1 SE/FSE^T2 FSE^STIR IR-FSE^T2 FLAIR IR-FSE^GRE^DWI^ |TR|400-700|3000+|3000+|9000+|VAR|8000+| |TE|MIN; <30|85-120|40-70|90-120|In phase|MIN; ~70-100| |ETL|3-5 If FSE|16-32|12-24|16-28|N/A|N/A| |BANDWIDTH|15khz/31khz|20khz|20khz|20khz|VAR|63khz+| |FREQ (pixel)|.5 - .8mm|.5 - .8mm|.8 - 1.2mm|.8 - 1mm|1mm|1.3mm| |PHASE (pixel)|.5 - .8mm|.5 - .8mm|.8 - 1.2mm|.8 - 1mm|1mm|1.3mm| |NEX|2 - 4|2 - 4|2 - 4|2 - 4|2 - 4|4 - 12| |SLICE THICKNESS|2 - 3mm|2 - 3mm|2 - 3mm|2 - 3mm|2 - 3mm|2 - 5mm| |SLICE SPACING|10%|10%|10%|10%|10%|10%| |NOTES| -Pre and post contrast \\ -Choose FSE for metal artifacts| -High ETL may cause blurring \\ -If using FatSat, decrease TE to 70-80 -TE 500+ for myelo| \\ -Superior FatSat over large FOV \\ -Do not scan post contrast -Low inherent SNR| \\ -Sensitive to subtle contrast  \\ -Scan with 2 concat \\ -Keep TR very long|Many variants \\ -T1: FLASH;VIBE \\ -T2*:GRE;MEDIC \\ -T2* for Hemo \\ -Can be 3D for thin slice \\ -Can be 3D for isotropic|-Low resolution \\ -Low SNR \\ -Not sensitive to gad \\ -Scan for ischemia \\ -EPI artifacts| Provided by Scott Johnson, MS RT (R)(MR) - edited by meuon, a weird spinning particle in the machine.