MRI Standard Parameters

ParameterT1 SE/FSET2 FSESTIR IR-FSET2 FLAIR IR-FSEGREDWI
TR400-7003000+3000+9000+VAR8000+
TEMIN; <3085-12040-7090-120In phaseMIN; ~70-100
ETL3-5 If FSE16-3212-2416-28N/AN/A
BANDWIDTH15khz/31khz20khz20khz20khzVAR63khz+
FREQ (pixel).5 - .8mm.5 - .8mm.8 - 1.2mm.8 - 1mm1mm1.3mm
PHASE (pixel).5 - .8mm.5 - .8mm.8 - 1.2mm.8 - 1mm1mm1.3mm
NEX2 - 42 - 42 - 42 - 42 - 44 - 12
SLICE THICKNESS2 - 3mm2 - 3mm2 - 3mm2 - 3mm2 - 3mm2 - 5mm
SLICE SPACING10%10%10%10%10%10%
NOTES -Pre and post contrast
-Choose FSE for metal artifacts
-High ETL may cause blurring
-If using FatSat, decrease TE to 70-80 -TE 500+ for myelo

-Superior FatSat over large FOV
-Do not scan post contrast -Low inherent SNR

-Sensitive to subtle contrast 
-Scan with 2 concat
-Keep TR very long
Many variants
-T1: FLASH;VIBE
-T2*:GRE;MEDIC
-T2* for Hemo
-Can be 3D for thin slice
-Can be 3D for isotropic
-Low resolution
-Low SNR
-Not sensitive to gad
-Scan for ischemia
-EPI artifacts