MRI Parameter Relationships

This is an old revision of the document!


MRI Paramter Relationships

Parameter IncreasedTIMESNRCNRRESOLUTIONARTIFACTS AND NOTES
TR++-N/AIncreased TR reduces T1 weighting
TEN/A-+N/AIncreased TE increases T2 weighting
ETL--+N/A High ETL can cause image blur; increases T2 weighting; MT effects; reduces susceptibility
BANDWIDTH--N/AReduces echo spacing; increase image sharpness; decrease chem shift; reduce motion
FREQ MATRIX+-N/A+Increases TE
PHASE MATRIX+-N/A+
FOVN/A++N/A-Reduces wrap artifact
NEX+++N/AN/AReduces motion; SNR increase sqrt2
SLICE THICKNESSN/A++N/A-Increases volume averaging

Provided by Scott Johnson, MS RT (R)(MR) - edited by meuon, a weird spinning particle in the machine.

library/mri_parameter_relationships.1737495485.txt.gz · Last modified: 2025/01/21 21:38
CC Attribution 4.0 International Except where otherwise noted, content on this wiki is licensed under the following license: CC Attribution 4.0 International