MRI Standard Parameters
| Parameter | T1 SE/FSE | T2 FSE | STIR IR-FSE | T2 FLAIR IR-FSE | GRE | DWI | 
|---|---|---|---|---|---|---|
| TR | 400-700 | 3000+ (1500+ if Restore Pulse) | 3000+ | 9000+ | VAR | 8000+ | 
| TE | MIN; <30 | 85-120 | 40-70 | 90-120 | In phase | MIN; ~70-100 | 
| ETL | 3-5 If FSE | 16-24 | 12-16 | 16-28 | N/A | N/A | 
| BANDWIDTH | 15khz/31khz | 20khz | 20khz | 20khz | VAR | 63khz+ | 
| FREQ (pixel) | .5 - .8mm | .5 - .8mm | .8 - 1.2mm | .8 - 1mm | 1mm | 1.3mm | 
| PHASE (pixel) | .5 - .8mm | .5 - .8mm | .8 - 1.2mm | .8 - 1mm | 1mm | 1.3mm | 
| NEX/Averages | 2 - 4 | 2 - 4 | 2 - 4 | 2 - 4 | 2 - 4 | 4 - 12 | 
| SLICE THICKNESS | 2 - 3mm | 2 - 3mm | 2 - 3mm | 2 - 3mm | 2 - 3mm | 2 - 5mm | 
| SLICE SPACING | 10% | 10% | 10% | 10% | 10% | 10% | 
| NOTES |  -Pre and post contrast  -Choose FSE for metal artifacts  |  -High ETL may cause blurring  -If using FatSat, decrease TE to 70-80 -TE 500+ for myelo  |   TI 150-180ms for 1.5T -Superior FatSat over large FOV -May suppress lesions if scanned post contrast -Low inherent SNR  |   -Sensitive to subtle contrast -Scan with 2 concat -Keep TR very long  | Many variants  -T1: FLASH;VIBE -T2*:GRE;MEDIC -T2* for Hemo -Can be 3D for thin slice -Can be 3D for isotropic  | -Low resolution  -Low SNR -Not sensitive to gad -Scan for ischemia -EPI artifacts  | 
	
Provided by Scott Johnson, MS RT (R)(MR) - edited by meuon, a weird spinning particle in the machine.